Commissioning
PR5230 Instrument Manual
EN-126
Sartorius
5.7.3
Configuring Limit Values
Each limit value consists of a switch-on and a switch-off point for definition of a hysteresis. The three pairs of
values must be entered according to the same principle. The limit values always refer to the gross weight.
Values between -0.01 x Max and 1.01 x Max of the relevant scale are permissible.
Define the limit values parameters under
-[Limit parameter].
Press
to select the specific parameter.
Press
to select the respective values.
Setup/
Limit parameter
Determine the limit values
Limit 1 on
890
kg
Action -no
action-
Limit 1 off
900 kg
Action -no
action-
Limit 2 on
300 kg
Action -no
action-
Limit 2 off
290 kg
Action -no
action-
Setup/
Digital I/O parameter
Assign the outputs
Output 1
limit 1 out
X16=1
Output 2
limit 2 out
X17=1