PR5230 Instrument Manual
Options
Sartorius
EN-35
3.2.10
Technical Data
Note:
Further technical data of PR5230 see Chapter 14.
3.2.10.1
General Data
Accuracy
<10000 e (Klasse III) according to OIML R76/EN 45501
Connection for the non Ex area via 4mm
2
screw terminals
Connection for the Ex area
via 4mm
2
screw terminals
Intrinsically safe output
Protection class Ex ib IIC
Potential equalization
Via screw terminal PE, see Chapter 3.2.8.4
3.2.10.2
Power Connection 230 V AC
Supply voltage
100…230 V AC ±10 %
Power consumption
16 VA
Frequency
50…60 Hz ±2 Hz
3.2.10.3
Power Connection 24 V DC
Supply voltage
24 V DC ±10 %
Power consumption
12 W
3.2.10.4
Output Parameter (intrinsically safe) for the Connection of Load Cells/Junction
Box/Weighing Platforms
The connection will be done via 7-pin plug (marked with ‚X100’ and ‚Exib’) on the weighing electronic board.
Uo
21,0 V between the lines
lo 143
mA
Po 1,55
W
Co
188 nF between the lines
Lo 0,12
mH