Global SPM Variables
PR5230 Instrument Manual
EN-176
Sartorius
10
Global SPM Variables
For communication via OPC, the following variables are available:
BOOL
R
Digital output 1
X8
BOOL
R
Digital output 2
X9
BOOL
R
Digital output 3
X10
BOOL
R
Digital input 1
X0
BOOL
R
Digital input 2
X1
BOOL
R
Digital input 3
X2
BOOL
R
Output limit 1
X16
BOOL
R
Output limit 2
X17
BOOL
R
Output limit 3
X18
BOOL R ADC
error
X32
BOOL
R
Above Max
X33
BOOL
R
Overload, above (Max + plus the range that is still permissible)
X34
BOOL R Below
zero
X35
BOOL
R
Zero +/- 1/4 d
X36
BOOL
R
Within the zero setting range
X37
BOOL
R
The weight is stable
X38
BOOL
R
Below zero or above Max
X39
BOOL
R
Load cell measuring signal negative
X40
BOOL
R
Load cell measuring signal > 36 mV
X41
BOOL
R
Internal arithmetic error; CAL data are perhaps faulty
X42
BOOL
R
ADC error, overload
X43
BOOL R Command
error
X48
BOOL R Command
active
X49
BOOL
R
Power fail signal
X50
BOOL
R
ADC test active
X56
BOOL R Calibration
active
X57
BOOL
R
Instrument is tared
X58
BOOL
R/W
Read / write marker bit 1
X64
BOOL
R/W
Read / write marker bit 2
X65
BOOL
R/W
Read / write marker bit 3
X66
BOOL
W
Switch D11 to net weight
X72
BOOL
W
Set the instrument to zero
X112
BOOL
W
Set tare of the instrument
X113
BOOL
W
Reset the tare of the instrument
X114
BOOL
W
Start the ADC test
X115
BOOL
W
Finish the ADC test
X116
BOOL
W
Reset the power fail signal
X117
BOOL
W
Set the fixtare value as tare
X118
BOOL
W
Store the current gross weight in the fixtare memory (D31)
X119
BOOL W Start
printing
X120
BOOL W Reset
error
X121