PR5230 Instrument Manual
Commissioning
Sartorius
EN-83
That means:
-
'DHCP' is activated.
-
‘Host name’ is initialized e.g. PR5230-
8BB499
(instrument type-
MAC ID
).
Example of MAC ID: 00-aa-00-
8b
-
b4
-
99
This ensures that a valid address for identification of the instrument in the network can be assigned to the
instrument, see Chapter 5.6.6.
Note:
The last 3 bytes of the MAC ID are displayed. A label with the complete MAC ID is fitted to the
outside of the device.
An device set to ’active’ DHCP (default/factory setting) and connected to an IT network
(company network) with a DHCP server does not require further actions except for a
2...3-minute waiting time. Subsequently, a network connection is established automatically
(device <-> workstation/PC).
Temporarily connected PCs must have the following network adaptor properties (DHCP/DNS automatic):
PR5230 can be operated only, if a PC is connected!
For commissioning, the first network contact is possible only by finding the IP address/subnet mask under
DHCP ’active’ (factory setting) automatically.