Asycom Protocol
PR5230 Instrument Manual
EN-162
Sartorius
8
Asycom Protocol
This function enables the use of the following commands of the old EW-command set, e.g. PR1612- command
set.
No license is required for using.
8.1
Commands of the Weight Function
Command Reply
Description
WGA
QGAwwwwwwemz
Read gross weight
WNA
QNAwwwwwwemz
Read net weight
WTA
QTAwwwwwwemz
Read tare weight
wwwwwwe = Weight with sign and exponent
(e.g. '002340' = 002.34 g; 001005 = 100 kg)
e = Exponent
m = 1: no motion/Stand still
z = 1: tared with status request
Possible error messages:
E100@0 general error, for @ see Chapter 8.3
E30000 BCC error
E50000 analog test active
E60000 weighing point unknown
WZA
Q
Set instrument to zero
Possible error messages:
E100@0 general error, for @ see Chapter 8.3
E30000 BCC error
E60000 weighing point unknown
E70000 weighing point tared or no standstill
WSA Q
Set
tare
WSAwwwwwwe Q
Fixtare
Possible error messages:
E100@0 general error, for @ see Chapter 8.3
E30000 BCC error
E60000 weighing point unknown
E63000 Fixtare > MAX
E70000 weighing point tared or no standstill
WFA Q
Reset
tare
Possible error messages:
E100@0 general error, for @ see Chapter 8.3
E60000 weighing point unknown
E70000 weighing point tared or no standstill
ZS1Awwwwwwewwwwwwe QZS1A
Set Limit 1
ZS2Awwwwwwewwwwwwe QZS2A
Set Limit 2
ZS3Awwwwwwewwwwwwe QZS3A
Set Limit 3
Possible error messages:
E20000 error in telegram
E60000 weighing point unknown
E63000 Fixtare >MAX
I
QIwwwww ddssnnnnn
d = Unit, variable (mg, g, kg, t, lb, oz)
ss = Scale interval (01, 02, 05, 10, 20, 50)
nnnnn = Span in mV/V (01234 acc. 0.1234 mV/V)