Commissioning
PR5230 Instrument Manual
EN-118
Sartorius
[Date
&
Time]
The date and time are printed as DD.MM.YYYY HH:MM:SS
Example for printing
PrintlayoutItem1
Date
&
Time
PrintlayoutItem2
Sequencenumber
PrintlayoutItem3
Grossweight
PrintlayoutItem4
Net weight
PrintlayoutItem5
CR/LF
PrintlayoutItem6
-none-
5.6.5
Fieldbus parameter
This menu item can only be selected if a fieldbus card is installed in slot 2.
Which protocol is displayed automatically depends on the Fieldbus card installed:
[ProfiBus-DP] for PR1721/41, [InterBus-S] for PR1721/42, [DeviceNet] for PR1721/44, [ProfiNet I/O] for
PR1721/46 and [EtherNet-IP] for PR1721/47.
Define the fieldbus parameters under
-[
Fieldbus parameter
].
Which additional parameters are required, depends on the interface type.
Press
to select the specific parameter.
Press
to select the respective values.
[ProfiBus-DP]
Setup/
Fieldbus parameter
fieldbus protocol
ProfiBus-DP
ProfiBus-DP address
1
Enter address
[InterBus-S]
Setup/
Fieldbus parameter
Fieldbus protocol
InterBus-S
[DeviceNet]
Setup/
Fieldbus parameter
Fieldbus protocol
DeviceNet
DeviceNet baudrate
500k
Select 500, 250 or 125 k
DeviceNet MAC-ID
1
Enter address 1 ... 62
[CC-Link]
Setup/
Fieldbus parameter
Fieldbus protocol
CC-Link
CC-Link baudrate
156k
Select 156k, 625k, 2.5M, 5M or 10M
CC-Link ID
1
Enter address