PR5230 Instrument Manual
Commissioning
Sartorius
EN-107
5.4.12
Parameter Input
Open the menu via
-[Calibration]-[Param].
WP A/
Calibration
Measuretime
160 ms
Digital filter
bessel
Fcut 2.00
Hz
This line is shown only, if the filter is on.
External supply
Above 8V
Test mode
absolute
W
&
M
none
Standstill time
0.50 s
[Measuretime]
Select the measurement time; possible values: 5, 10, 20, 40, 80, 160, 320, 640, 960, 1280, 1600 ms. In 'legal-
for-trade' mode select
1 s.
[Digital filter]
The digital filter can be switched on only with the [measurement time] set to
160 ms.
Select the filter characteristic: [off], [Bessel], [aperiod.], [Butterw.], [Tscheby.].
[Fcut]
Enter the cut-off frequency for the low-pass filter (0.1…80 Hz); the setting is dependent on the measurement
time.
[External supply]
The external supply voltage is selected. Possible values: >8 V and
8 V.
[Test mode]
With [absolute], the test value is determined when the test is called. With [relative], the deviation from the
initially stored test value is displayed; see Chapter 5.4.10.
[W
&
M]
Setting for 'legal-for-trade' mode; select [none], [OIML], [NTEP] (for USA) or [NSC] (for Australia); see
Chapter 5.4.12.1.
[Standstill time]
Parameters [Standstill range] and [Standstill time] are required for determining the mechanical stability of the
scale. Input in seconds; permissible range: 0.01…2 sec. If 0 is set, stability is not checked. The stability time
must not be less than the measurement time.