Installing the Instrument and Plug-in Cards
PR5230 Instrument Manual
EN-60
Sartorius
4.3.4.1
Status-LEDs on Fieldbus Card
LED Meaning
Frequency Color
Flashing
1 Hz green
Fieldbus is initialized and ready for operation.
Flashing
2 Hz green
Fieldbus is not initialized.
Flashing
1 Hz red
Check error in ASIC and FLASH ROM, fieldbus is defect.
4.3.4.2
Status Display
Requirements
In
-[Display items] the items are configured, see Chapter 5.6.7.
In
-[HW-Slots] PR1721/41 is selected.
Display
1 2 3 4
LED --- --- --- ---
1
2 3 4
Lit
-
-
-
:
no function
Lit
grn
:
Fieldbus is online, data
transmission is possible.
Flashing 1 Hz
red
:
Input/output length
configuration error.
Lit
red
:
Fieldbus is off-line, data
transmission is not possible.
Flashing 2 Hz
red
:
User parameter error.
Flashing 4 Hz
red
:
Error in ASIC
Legend
--- off
red red
grn green
4.3.4.3
Connecting Diagram for a Master with three Slaves
Caution!
The cable screens must be connected in the metal sleeves.
Don’t take the screens into the unit!
Before, during and after installation, make sure that the sealing rings are seated
correctly.