PR5230 Instrument Manual
Commissioning
Sartorius
EN-101
Possible error messages
The dead load entered in mV/V plus maximum capacity in mV/V is higher
than 3 mV/V (= 36 mV).
This message displays, if the scale is not stable.
Remedial action
Check the mechanical function of the scale.
Adapt the filter setting; reduce the resolution, if necessary.
Adapt the stability conditions.
This message displays, if the Measurement signal is negative when determining
the dead load with [by load].
Cause
Load cells connected with wrong polarity or defective.
This message displays, if dead load entered in mV/V is higher than 5mV/V.