Commissioning
PR5230 Instrument Manual
EN-92
Sartorius
5.3.4
Function INFO
When you press
the program releases and status messages are displayed. The
key also has other
functions; see Chapters 5.4.1.1 and 5.4.9.
Info
Show version
Show status
Show HW-slots
When you select [Show version], the installed program releases and the board number are displayed:
Info/
Version
Firmware Rel.
00.90.00.55992
Firmware release and creation date
2009-08-04-14:21
Application Rel.
01.00.00
Application release and creation date
2009-06-09-14:16
Bios
Rel. 00.98.00. 55992
BIOS release and creation date
2009-08-04-14:18
Boardnumber 327923612
Main board identification number
(different from the device serial number)
When you select [Show status], instrument status information is displayed:
Info/
Status
Free system RAM
5328 of 15204 kb
CAL-Switch opened
opened = CAL switch is open,
closed = CAL switch is closed.
Builtin RS232
no signal
When you select [Show HW-slots], the installed plug-in cards are displayed:
Info/
HW-Slots
Builtin
RS485
Firmly installed serial interface
Builtin RS232
Firmly installed serial interface
Builtin analog
out
Optional interface (Slot 1, B1*)
Builtin digital
i/o
Firmly installed digital interface
PR1721/41 ProfiBus-DP
Only field bus cards (Slot 2, B2*)
PR5230/xx
ADC
For weighing electronics board (W1 or WE1)
When selecting ‚PR5230/xx’, factory serial number and production date are displayed:
Serial No 30252 00055
Prod.date: 2010-04-26 13:48:30
5.3.5
Setup Function (VNC)
Press
to configure the main operating parameters. The configuration depends on the application licenses
registered and the plug-in cards installed.
Calibration is in a simple dialogue. Compliance with the relevant (verification) standards must be checked by
the person commissioning the instrument or the verification officer. To protect the calibration data from
overwriting, close the corresponding CAL switch on the weighing electronic board. On legal-for trade
instruments, the CAL switch must be sealed in the closed (write-protected) position; see Chapter 5.1.1.