PR5230 Instrument Manual
Specifications
Sartorius
EN-183
14.4
Effect of Ambient Conditions
14.4.1
Environmental Conditions
Temperature range
Ambient temperature for operation
-10...+40 °C
Ambient temperature 'not legal for trade'
-10...+50 °C
Power-on
temperature
0...+40
°C
Storage/transport
-20...+70
°C
Humidity
<95 %, no condensation, (acc. to IEC 68-2)
Protection type
IP 66 housing
Height <2000
m
14.4.2
Electromagnetic Compatibility (EMC)
All data in compliance with NAMUR NE 21, EN 45501 and EN 61326
Housing
Radio frequency electromagnetic fields
(80 – 3000 MHz)
EN 61000-4-3
10 V/m
Electrostatic discharge (ESD)
EN 61000-4-2
6/8 kV
Signal and control
lines
Fast transients (burst)
EN 61000-4-4
3 kV
Peak voltages (surge) 1.2/50
μ
s
EN 61000-4-5
2 kV
Conducted disturbances by radio frequency
(0.15 – 80 MHz)
EN 61000-4-6
10 V
Mains inputs
Fast transients (burst)
EN 61000-4-4
3 kV
Peak voltages (surge) 1.2/50
μ
s
EN 61000-4-5
1/2 kV
Conducted disturbances by radio frequency
(0.15 – 80 MHz)
EN 61000-4-6
10 V
Voltage dips
EN 61000-4-11
0/40/70 %
20/200/500 ms
Short interruptions
EN 61000-4-11
5 s
14.4.3
RF Interference Suppression
Electromagnetic emission
In acc. with EN 61326, limit value class A