Fieldbus Interface
PR5230 Instrument Manual
EN-172
Sartorius
9.3
Fieldbus Register
9.3.1
Register 0: IO Status Bits for Reading
(Dynamic status), only reading is permitted
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Input 3
Input 2
Input 1
Byte 1
Output 3 Output 2
Output 1
Byte 2
Limit 3
Limit 2
Limit 1
Byte 3
9.3.2
Register 1: Scale Status
(Dynamic status), only reading is permitted
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
DIM
STND
INZSR
CZERO
BELOW0
OVL
>Max
ADUERR
Byte 1
E6
E1
E3
E7
Byte 2
PowerFail ActionActive CmdError
Byte 3
TareActive CalChanged TestActive
Byte 0 corresponds to byte 7 in the output area; for weight error, see also the table in Chapter 5.5.1.
ADUERR
Error of analog conversion/ load cell circuit (OR function of the E1,E3,E7 bits)
>Max
Gross value is higher than Max (maximum capacity); the full scale deflection was
exceeded.
OVL
Scale overload; Gross > Max + overload,
Error 2
BELOW0
Gross weight is negative (< 0-1/4 d)
CZERO
Center of zero; weight within
0 +/- 1/4 d
INZSR
The gross weight is within the zero setting range
STND
The scale is stable
DIM
The gross weight has exceeded the full scale deflection (0-1/4 d > weight value >
Max+overload), (OR function of bits
BELOW0, OVL
).
E7
The input signal is negative (inverse conversion)
,
Error 7
E6
No sense voltage, or Sense voltage is low,
Error 6
E3
The input signal is > 36mV (no end of conversion),
Error 3
E1
Arithmetic error (overflow),
Error 1
CmdError
Error during execution (
cmdError
); e.g. the 'taring' operation is not handled, because the
scale is not stable. The error is stored in
LastError
(register 4). The bit is reset with the
ResetError
bit (register 2).
ActionActive
The operation is handled; handling is busy
PowerFail
Power failure; is always set after power on. The PowerFail bit is reset with the
ResetPWF
bit (register 2) 'Reset power failure'.
TestActiv
The analog test is busy
TareActiv
The instrument was tared.