Installing the Instrument and Plug-in Cards
PR5230 Instrument Manual
EN-64
Sartorius
4.3.6.1
Status-LEDs on Fieldbus Card
LED Meaning
Frequency Color
Flashing
1 Hz green
Fieldbus is initialized and ready for operation.
Flashing
2 Hz green
Fieldbus is not initialized.
Flashing
1 Hz red
Check error in ASIC and FLASH ROM, fieldbus is defect.
4.3.6.2
Status Display
Requirements
In
-[Display items] the items are configured, see Chapter 5.6.7.
In
-[HW-Slots] PR1721/44 is selected.
Display
1 2 3 4
LED --- --- --- ---
1 2
3
4
Lit
-
-
-
:
No function.
Lit
-
-
-
:
Not powered, not online.
Lit
-
-
-
:
No function.
Lit
-
-
-
:
No power to device.
Lit
grn
:
Link OK, online, connected.
Lit
grn
:
Device operational.
Flashing
grn
:
Online, not connected.
Flashing
grn
:
Data size > configuration.
Lit
red
:
Critical link error.
Lit
red
:
Unrecoverable error.
Flashing
red
:
Connection timeout.
Flashing
red
:
Minor error.
Legend
--- off
red red
grn green