Please read this user’s manual thoroughly to ensure correct use of the 
instrument.  Carefully store this manual for future reference. 

 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

  1.  Cautionary Notes for Safe Use of the Products

The following symbol is indicated on the instrument and in this manual to 
ensure safe use. 

 

 

This symbol is indicated where the operator must refer to instructions 
in the manual in order to protect both personnel and the instrument. 
This symbol on the product indicates that the operator must refer to 
an explanation in the user’s manual in order to avoid the risk of injury 
or death of personnel and damage to the instrument. The manual 
describes how the operator should exercise special care to avoid 
electric shock or other dangers that may in injury or loss of life.  

 

The purpose of the following conventional symbol in this manual is: 

NOTE 

 

Draws attention to information essential for understanding the 
operation and features. 

 

  2.  Handling Precautions and Installation Location 

N

 

OTE 

  Note that these current transformers are designed for use with 

general-purpose instruments and have not yet gained any official 
certification. They cannot therefore be used for the purpose of 
ascertaining wattage or other electric parameters. 

  Use each current transformers under a load no greater than its 

rated level. When calculating the load, you must calculate the 
sum of the load given by the instruments to be connected to the 
transformer and the load given by the lead-wire of the 
secondary-stage cable. 

  The inner-walls of the clamp are treated with rust preventive. But 

if rusted, remove the rust with CRC-5-56 (goods on the market) 
and spray rust preventive on the inner-walls again for recovery. 

 

Avoid installing the current transformers in locations subject to: 
ambient temperatures outside the specified range of -20 

°

C to 50 

°

C, where 

condensation is likely; ambient humidity exceeding 80 % RH, where 
condensation is likely; relatively large amounts of flammable or explosive 
gas, saline substance, lampblack or dust; splashes of water, oil or solvent; 
radioactive rays or intense electric or magnetic fields; frequent mechanical 
vibration or shock; or direct sunlight.  

 3. Installation 

W

 

ARNING 

Since there is a risk of electric shock during installation, be sure to 
observe the following. 

  Before connecting the transformer to any other equipment, make 

sure the circuit is not receiving any power in order to prevent 
electric shock. 

  Fasten the connecting terminals securely so they will not loosen 

readily. 

  Do not touch the terminals or any other part of the transformer 

when the circuit is live. 

  Do not use the transformer under a circuit current greater than 

the level specified for that model. 

  Do not use the transformer under a circuit voltage greater than 

500 V AC. 

 

 

N

 

OTE

The CTW contains a device for suppressing hazardous voltages that 

ay arise when the secondary line is open-circuited. 

m

 

Install the transformer referring to the following figures. 

How to Install and How to Remove

Example of Installation

Cable

Power-source side

K

Load side

L

Secondary-

stage cable

l side=black

k side=white

Banding band

B

A

How to Install

Push the B part until it clicks to lock

the A part.

How to Remove

Loosen the A part using a flat-blade

screw driver to unlock, and pull the B

part in the direction of the arrow to

release.

 

●   Attach the transformer to the cable according to the orientation of the 

K and L symbols, as shown in the figure.  

●   Be careful not to make a gap between the inner-walls of the clamp. 

●   Be careful not to trap any dust between the inner-walls of the clamp. 

●   Use  a  crimp-on  terminal  for  0.75  mm

2

  cables  to  connect  the 

secondary-stage cable. 

●   The wires of the secondary-stage cable are color-coded as follows.  

k side=white; l side=black 

●   Fix the transformer with a banding band or the like as shown in the 

figure above to prevent it from falling. 

 

IM 77C01W03-01E

 3rd 

Edition: 

Jan. 

2005

 

User’s 
Manual 

Cross-check the Model 

Make sure the model and suffix codes shown on the data plate attached to 
the transformer are the same as specified in your purchase order. 

Models Names 

CTW15 

Clamp-on Current Transformer (for 100 A / 1 A) 
Applicable cable diameter: less than 

Φ

24 mm 

CTW35 

Clamp-on Current Transformer (for 300 A / 1 A) 
Applicable cable diameter: less than 

Φ

36 mm 

 

Models  CTW15 and CTW35 

Clamp-on Current Transformers

IM 77C01W03-01E   3rd Edition : 2005.01.15-00 