background image

T5000 Series

Telaire

®

 Airestat

 CO

2

/

Temperature Transmitter

GE Measurement & Control Solutions offers high-
volume manufacturing capabilities for its Telaire 
products, a global sales force, and additional 
engineering resources to support your sensing 
application needs.

Telaire

®

 Technology

The technology is based on the absorption of 
light in a gold-plated, reflective light pipe or 
waveguide diffusion gas chamber. A gas permeable 
polytetrafluoroethylene  (PTFE) filter prevents 
particulate and water contamination of the 
sensor. Light is absorbed in proportion to the CO

2

 

concentration and the remaining light is measured 
and converted into an analog signal.

ABC Logic

 (Automatic Background Calibration)

The Telaire product line offers patented ABC Logic- 
software for self-correction of drift to better than 
±20 ppm per year. The system is virtually free of 
maintenance and typically has a lifetime of more 
than 10 years.

Features:

• Low-cost

• Reliable (15 years of low-cost infrared sensor 

manufacturing)

• Lifetime calibration guarantee (ABC Logic™ on)

• Pre-calibrated 

• Maintenance free (calibration maintained with 

patented ABC Logic)

• Flexible configurations (see matrix, page 2)

• Intuitive VLI display

• Three available outputs: analog 0 to 10V, 0 to 5V, 

passive temperature (20K thermistor), and digital 
UIP interface (RS-232 TTL level)

• User-definable thresholds

• RoHS and WEEE compliant

GE

Measurement & Control Solutions

Summary of Contents for T5000 Series

Page 1: ...ing light is measured and converted into an analog signal ABC Logic Automatic Background Calibration The Telaire product line offers patented ABC Logic software for self correction of drift to better than 20 ppm per year The system is virtually free of maintenance and typically has a lifetime of more than 10 years Features Low cost Reliable 15 years of low cost infrared sensor manufacturing Lifeti...

Page 2: ...utput For 0 5V output the part number requires 5V i e T5005 5V VLI Visual Limit Indicator GE Telaire Products introduces an exciting new alternative to traditional scientific LCD display of the CO2 value Our new T5000 Series transmitter with VLI Visual Level Indicator communicates intuitively to the average end user The correlation between a CO2 PPM reading and adequate ventilation may not be full...

Page 3: ...aximum accuracy Operating Conditions 32 F to 122 F 0 C to 50 C 0 to 95 RH non condensing Storage Conditions 4 F to 158 F 20 C to 70 C 0 to 95 RH non condensing Temperature NTC 20k thermistor Output Analog 0 to 10V or 0 to 5V Connections Screw terminals for 18 to 28 AWG Power Supply Requirements 18 to 30 VAC RMS 50 60Hz or 18 to 42 VDC polarity protected Power Consumption 1 75 VA average 2 75 VA pe...

Page 4: ...cations are subject to change without notice GE is a registered trademark of General Electric Company Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective companies which are not affiliated with GE 920 420D ...

Reviews: