38
Operating Instructions Cubis
®
MCE
System Settings
Parameters in the “Motion Sensor” Submenu
Parameters
Setting values
Explanation
CONTROL
OFF
Deactivates the motion sensor:
No
gestures are recognized.
ON*
Activates the motion sensor.
NUM.GEST.
2 GESTUR.*
Activates the control of functions with 2 gestures. Gestures: Left, right
4 GESTUR.
Activates the control of functions with 4 gestures. Gestures: Left, right, top,
bottom
LEFT CMD.
PRINT*
Left command of the gesture control corresponds to the [Print] key.
ZERO
Left command of the gesture control corresponds to the [Zero] key.
TARE
Left command of the gesture control corresponds to the [Tare] key.
NONE
No
function is assigned to the left command of the gesture control.
RIGHT.CMD.
PRINT*
Right command of the gesture control corresponds to the [Print] key.
ZERO
Right command of the gesture control corresponds to the [Zero] key.
TARE
Right command of the gesture control corresponds to the [Tare] key.
NONE
No
function is assigned to the right command of the gesture control.
UP.CMD.
PRINT*
Upper command of the gesture control corresponds to the [Print] key.
ZERO
Upper command of the gesture control corresponds to the [Zero] key.
TARE
Upper command of the gesture control corresponds to the [Tare] key.
NONE
No
function is assigned to the upper command of the gesture control.
DOWN.CMD.
PRINT*
Lower command of the gesture control corresponds to the [Print] key.
ZERO
Lower command of the gesture control corresponds to the [Zero] key.
TARE
Lower command of the gesture control corresponds to the [Tare] key.
NONE
No
function is assigned to the lower command of the gesture control.
* Factory setting
Parameters in the “COM-RS232” Submenu
Parameters
Setting values
Explanations
DAT.REC.
SBI.WEIGH.
Enables a faster SBI communication, only for software commands to the
weighing module.
XBPI
Extended range of commands to control numerous balance functions with
binary protocol for direct communication with the weighing module.
SBI*
−
Enables SBI communication.
−
The data is output to a PC or control unit. Enables the use of ESC commands
from a PC to control the basic balance functions with ASCII protocol.
OFF
Deactivates the automatic data output.
* Factory setting