Operating Instructions Cubis
®
MCE
63
Malfunctions
Fault
Cause
Remedy
Chapter,
page
For a conformity-
assessed device: The
[
Invalid
weight value]
display appears.
The cause of this fault is displayed after
pressing the [Invalid value] key. At the
same time, the “Toggle between weight
units” function is locked.
−
LEVEL.ERR.: The device must be
leveled.
Level the device.
−
ISOCAL.ER.: The device must be
calibrated.
Calibrate the device.
−
WARMUP.xx.: The device is in the
warm-up phase and has not yet
reached its operating temperature.
xx = remaining time in minutes
Comply with the warm-up time
after switching the device on.
−
VALUE.ERR.: The displayed value is
invalid.
Set the device to zero.