Operating Instructions Cubis
®
MCE
51
Operation
Procedure
t
If the isoCAL auto-start function is set (configuration values for the
“ISOCAL” parameter, see Chapter “7.3.1 Parameters in the “Setup” Main
Menu”, page 34):
y
If the isoCAL function is triggered according to the aforementioned
conditions: The [isoCAL] button flashes in the operating display.
y
In the operating display, a time display counts down from 15 seconds
to 0.
y
If
no
load change or
no
operation takes place on the device before
the expiration of the time display: The isoCAL function starts.
t
If the isoCAL manual start function is set (configuration values for the
“ISOCAL” parameter, see Chapter “7.3.1 Parameters in the “Setup” Main
Menu”, page 34):
y
If the isoCAL function is triggered: The [isoCAL] button flashes in the
operating display.
t
Press the [isoCAL] button.
y
The isoCAL function starts.
8.7 Internally Calibrating and Adjusting
the Device
Calibration and Subsequent Adjustment
Requirements
−
The weighing pan is unloaded.
−
The operating display shows a stable weight value.
Procedure
t
To set the internal adjustment in the menu: In the “SETUP” / “BALANCE”
menu, for the “CAL./ADJ.” parameter, select the “INT.CAL.”
configuration value.
t
To set the sequence of the adjustment function to calibration and
subsequent automatic adjustment: In the “SETUP” / “BALANCE” menu,
for the “CAL.SEQ.” parameter, select the “ADJUST” configuration value.
t
Exit the menu.
t
Press the [Zero] key.
t
To start the internal calibration | adjustment: Press the [Adjust] button.
y
The internal calibration weight is loaded.
y
The [CAL.RUN.] display appears in the operating display during
calibration.
y
The device is automatically calibrated and the internal calibration weight
is removed.
y
The [CAL.END] display indicates the end of calibration.