Operating Instructions Cubis
®
MCE
39
System Settings
Parameters
Setting values
Explanations
BAUD
600
Sets the baud rate to 600 baud.
1200
Sets the baud rate to 1200 baud.
2400
Sets the baud rate to 2400 baud.
4800
Sets the baud rate to 4800 baud.
9600*
Sets the baud rate to 9600 baud.
19200
Sets the baud rate to 19200 baud.
38400
Sets the baud rate to 38400 baud.
57600
Sets the baud rate to 57600 baud.
115200
Sets the baud rate to 115200 baud.
PARITY
ODD*
Applies an odd parity.
EVEN
Applies an even parity.
NONE
Does not apply a parity.
HANDSHK.
SOFTWARE
Sets the handshake protocol to software handshake.
HARDWARE*
Sets the handshake protocol to hardware handshake.
NONE
Does
not
set a handshake protocol.
DATABIT
7 BITS
Sets the number of data bits to 7.
8 BITS*
Sets the number of data bits to 8.
* Factory setting
Parameters in the “PC-USB” Submenu
Parameters
Setting values
Explanations
DAT.REC.
SBI.WEIGH.
Enables a faster SBI communication, only for software commands to the
weighing module.
XBPI
Extended range of commands to control numerous balance functions with
binary protocol for direct communication with the weighing module.
SBI*
−
Enables SBI communication. The data is output to a PC or control unit.
−
Enables the use of ESC commands from a PC to control the basic balance
functions with ASCII protocol.
PC.SPREAD.
Enables data output to a spreadsheet program via a direct PC connection.
PC.TEXT
Enables a direct PC connection to be established to a text processing program.
OFF
Deactivates the data output.
* Factory setting