t
Place the indicated calibration weight on the balance. The calibration
can only be started, if the calibration weight used is within the defined
limits.
y
The [+] sign is shown if the applied weight is too high.
y
The [-] sign is shown if the applied weight is too low.
t
Press the [Confirm] button on the keypad. This starts the calibration.
y
The [CAL.END] display indicates the end of calibration.
t
Remove the calibration weight after calibration is complete.
8.9 Printing Results
The results of the calibration can be printed in a GLP printout.
Procedure
t
Set the GLP printout in the menu (setting parameters, see Chapter “Pa
rameters in the “Settings for Printouts” Submenu”, page 41):
t
Calibrate the device.
y
Once the calibration is complete:
y
The printing process starts.
y
The GLP footer is printed.
8.10 Weighing and Printing with ID Marking
The device and every lot can be assigned an ID number. The ID numbers are
provided in the GLP header during GLP-compliant printing.
Requirements
− The ID marking is entered in the menu, e.g. “BATCH1” (see Chapter “7.3.5
Parameters in the “Input” Main Menu”, page 45).
−
The printing of the lot ID line in the GLP printout is activated in the menu
(see Chapter “7.3.5 Parameters in the “Input” Main Menu”, page 45).
− The GLP-compliant printout is activated (see Chapter “7.3.5 Parameters
in the “Input” Main Menu”, page 45).
− The format for the printout is set to 22 characters per line (configuration
parameters, see Chapter “7.3.3 Parameters in the “Data Output” Main
Menu”, page 41).