52
Operating Instructions Cubis
®
MCE
Operation
Manually Starting Adjustment after Calibration
Requirements
−
The weighing pan is unloaded.
−
The operating display shows a stable weight value.
Procedure
t
To set the internal adjustment in the menu: In the “SETUP” / “BALANCE”
menu, for the “CAL./ADJ.” parameter, select the “INT.CAL.”
configuration value.
t
To set the sequence of the adjustment function to calibration without
subsequent automatic adjustment: In the “SETUP” / “BALANCE” menu,
for the “CAL.SEQ.” parameter, select the “CAL/ADJUST” configuration
value.
t
Exit the menu.
t
Press the [Zero] key.
t
To start the internal calibration: Press the [Adjust] button.
y
The internal calibration weight is loaded.
y
The [CAL.RUN.] display appears in the operating display during
calibration.
y
The calibration weight value is shown.
t
To start adjustment: Press the [Confirm] button on the keypad.
y
The device is calibrated and the internal calibration weight is removed.
y
The [CAL.END] display indicates the end of calibration.
8.8 Externally Calibrating and Adjusting
the Device (Not for Conformity-assessed
Models)
External Calibration with Factory Set Weight Value
Procedure
t
In the “SETUP” / “BALANCE” menu, for the “CAL./ADJ.” parameter,
select the “EXT.CAL.” configuration value.
Set the User-defined Weight Value
A weight value can be set for the calibration weight. This user-defined
weight value is used for the external calibration.
If no user-defined weight value is set: The preset weight value is used.
Procedure
t
In the “INPUT” menu, select the “CAL.WT.” configuration value.
t
Enter and confirm the desired weight value.
t
To use the user-defined weight value for the next calibration: In the
“Balance” menu, for the “CAL./ADJ.” parameter, select the “E.CAL.USR.”
configuration value.