70
Operating Instructions Cubis
®
MCE
Technical Data
15.7 Metrological Data
15.7.1 Models MCE32202P | MCE70201S | MCE50201S | MCE36201S
MCE32202P MCE70201S MCE50201S MCE36201S
Unit
Value
Value
Value
Value
Scale interval (d)
g
0.01 | 0.1
0.1
0.1
0.1
Maximum capacity (Max)
g
4200 | 32200 70200
50200
36200
Repeatability at 5% load
Standard deviation of the load values,
tolerance
g
0.04
0.1
0.1
0.1
Standard deviation of the load values,
typical value
g
0.01
0.04
0.04
0.02
Repeatability at approx. maximum
capacity
Standard deviation of the load values,
tolerance
g
0.1
0.1
0.1
0.1
Standard deviation of the load values,
typical value
g
0.02
0.04
0.04
0.02
Linearity deviation
Tolerance
g
0.05 | 0.2
0.5
0.5
0.2
Typical value
g
0.02 | 0.1
0.15
0.15
0.1
Deviation when load is off-center,
positions according to OIML R76
Test weight
g
10000
20000
20000
10000
Tolerance
g
0.2
0.5
0.5
0.3
Typical value
g
0.1
0.3
0.3
0.2
Sensitivity drift b10°C and +30°C ppm/K
2
2
2
2
Tare-maximum capacity: Less than 100%
of maximum capacity
Accuracy class according to Directive
2014/31/EU
–
II
II
II
Verification scale interval (e) according to
Directive 2014/31/EU
g
–
1
1
1
Minimum load (Min) according to
Directive 2014/31/EU
g
–
5
5
5
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing
g
8.2
82
82
82
Typical minimum initial weighing
g
20
82
82
82
Typical stabilization time
s
2
1.5
1.5
1.5
Typical measurement time
s
2
1.5
1.5
2