Technical Data
15.7.2 Models MCE36201P | MCE20201S | MCE11201S | MCE70200S | MCE36200S
MCE
36201P
MCE
20201S
MCE
11201S
MCE
70200S
MCE
36200S
Unit
Value
Value
Value
Value
Value
Scale interval (d)
g
0.1 | 1
0.1
0.1
1
1
Maximum capacity (Max)
g
10200 | 36200 20200
11200
70200
36200
Repeatability at 5% load
Standard deviation of the load values,
tolerance
g
0.1
0.1
0.1
0.5
0.5
Standard deviation of the load values,
typical value
g
0.02
0.02
0.02
0.2
0.2
Repeatability at approx. maximum
capacity
Standard deviation of the load values,
tolerance
g
0.1
0.1
0.1
0.5
0.5
Standard deviation of the load values,
typical value
g
0.02
0.02
0.02
0.2
0.2
Linearity deviation
Tolerance
g
0.2
0.2
0.2
1
1
Typical value
g
0.1
0.06
0.06
0.2
0.2
Deviation when load is off-center,
positions according to OIML R76
Test weight
g
10000
5000
5000
20000
10000
Tolerance
g
0.3
0.3
0.3
1
1
Typical value
g
0.2
0.2
0.2
0.6
0.5
Sensitivity drift b10°C and
+30°C
ppm/K
2
2
2
3
3
Tare-maximum capacity: Less than
100 % of maximum capacity
Accuracy class according to Directive
2014/31/EU
II
II
II
II
II
Verification scale interval (e) according
to Directive 2014/31/EU
g
1
1
1
10
1
Minimum load (Min) according to
Directive 2014/31/EU
g
5
5
5
50
50
Minimum initial weighing according to
USP (United States Pharmacopeia),
Chap. 41
Optimum minimum initial weighing
g
82
82
82
820
820
Typical minimum initial weighing
g
82
82
82
820
820
Typical stabilization time
s
1.5
1.5
1.5
1
1
Typical measurement time
s
2
2
2
1.2
1.2