Operating Instructions Cubis
®
MCE
45
System Settings
Parameters in the “Statistics” Submenu
Parameters
Setting values
Explanation
COMP.PRT.
OFF
Deactivates the component printout.
ON*
Activates the component printout.
TAR./STAT.
OFF*
Deactivates automatic taring after the components are saved.
ON
Activates automatic taring after the components are saved.
* Factory setting
7.3.5 Parameters in the “Input” Main Menu
Parameters
Setting values
Explanation
ID NO.
Maximum of
8 characters
09-0, A-Z, - ,
space
Saves the entered ID number for the device.
LOT ID
OFF*
Deactivates the prompt of lot ID number during data output.
ON
−
Activates the prompt of the lot ID number during data output.
−
The LOT-ID can be entered in writing in the line.
DATE
Saves the entered date.
TIME
Saves the entered time.
USER.PASS.
Maximum of
8 characters
09-0, A-Z, - ,
space
Saves the entered user password.
DEL.PASS.
YES
Deletes the entered user password.
NO*
Does
not
delete the entered user password.
SERV.PASS.
Max 8
characters
09-0, A-Z, - ,
empty
Activates service mode.
CAL.WT.
Changes the calibration weight for the adjustment or calibration process with
the user-defined weight value.
INTERV.
0-9999
seconds
−
Changes the interval for automatic data output for parameters “DATA.
OUTP./COM.SBI /AUTO.CYCL./INTERV.” and “DATA.OUTP./PRNT.
PARA./ACTIVATE/INTERV.”.
− “0” activates the fastest output rate (generally 100 ms).
* Factory setting