72
Operating Instructions Cubis
®
MCE
Technical Data
15.8 Recommended Calibration Weight
MCE32202P MCE70201S MCE50201S MCE36201S
Unit
Value
Value
Value
Value
External test weight
kg
30
70
50
30
Recommended accuracy class
F1
F1
F1
F1
MCE36201P MCE20201S MCE11201S
MCE70200S
Unit
Value
Value
Value
Value
External test weight
kg
30
20
10
70
Recommended accuracy class
F1
F1
F1
F1
MCE36200S
Unit
Value
External test weight
kg
30
Recommended accuracy class
F1
15.9 isoCAL Function
15.9.1 Models MCE32202P | MCE70201S | MCE50201S | MCE70200S
Unit
Value
isoCAL is triggered by the following criteria:
In the event of a temperature change
K
2
After a time interval
h
12
After successful leveling
With conformity-assessed devices, isoCAL
cannot
be turned off.
15.9.2 Models MCE36201S | MCE36201P | MCE20201S | MCE11201S | MCE36200S
Unit
Value
isoCAL is triggered by the following criteria:
In the event of a temperature change
K
4
After a time interval
h
12
After successful leveling