Cleaning and Maintenance
9.3 Assembling and Connecting the Device
Procedure
t
Replace the weighing pan and the associated components on the
weighing module (see Chapter “5.6 Positioning the Weighing Pan and
Associated Components”, page 29).
t
Re-connect the device to the power supply (see Chapter “6.3 Connecting
9.4 Maintenance Schedule
Interval
Component
Action
Chapter, page
Monthly to every 2 years, depending
on the operating conditions
Entire system
Contact Sartorius Service
9.5 Performing a Software Update
A software update can be installed from a USB mass storage device using
the device’s USB-A connection.
Requirements
−
The device is switched on.
−
The software update is saved on a USB mass storage device.
Procedure
t
Download the software update from the Sartorius website onto the USB
mass storage device.
t
If this relates to a zip file: Unzip the software update on a USB stick.
t
Insert the USB mass storage device with the software update into the
device’s USB-A connection.
t
Access the configuration menu and select the [UPDATE] main menu.
t
In the [UPDATE] main menu, select the [EXECUTE?] parameter.
y
The [COPYING] display appears in the operating display.
y
The [REBOOT] display appears in the operating display and the device
restarts.
y
The [UNPACK] display appears in the operating display.
y
The [RESTART] display appears in the operating display and the device
restarts.
y
During the restart, the [CHECK.APC] display appears first. The
[PROG APC] display then appears in the operating display for about
2 minutes.
t
Do
not
switch off the device!
y
The device confirms successful installation with the [PROG OK] display.
y
All identified components are updated with the new firmware.
y
The device confirms the end of the installation process with the [FINISH]
display.
y
The device restarts.
y
The device is ready for use.