System Settings
Parameters
Setting values
Explanation
ST.DEL.
NONE
Sets the stability delay to “none”: The stability symbol is displayed after the
stability criterion is reached.
SHORT*
Sets the stability delay to “short”: The stability symbol only appears after
a short delay in order to provide a reliable result despite fluctuations.
MEDIUM
Sets the stability delay to “medium”: The stability symbol only appears after
a longer delay in order to provide a reliable result in case of higher fluctuations.
LONG
Sets the stability delay to “long”: The stability symbol only appears after a long
delay in order to balance out major instability.
ZERO/TAR.
W/O STB.
Without stability: The function of the [Zero] or [Tare] key is executed
immediately once the key is pressed.
W/STAB.*
With stability: The function of the [Zero] or [Tare] key is only executed after
stability is achieved.
AT STAB.
At stability: The function of the [Zero] or [Tare] key is executed if stability exists
when the key is pressed.
AUTOZER.
ON*
−
Activates automatic zeroing.
− The display is automatically set to zero in case of a deviation of 0 less than (X).
OFF
Deactivates automatic zeroing. Zeroing must be triggered with the [Zero] key.
UNIT
The availability of units may depend on national legislation and is therefore
country-specific.
GRAMS*
The device displays the weight in grams.
KILOGR.
The device displays the weight in kilograms (not for semi-microbalances and
microbalances).
CARATS
The device displays the weight in carats.
POUNDS
The device displays the weight in pounds (not for semi-microbalances and
microbalances).
OUNCES
The device displays the weight in ounces (not for microbalances).
TROY OZ.
The device displays the weight in troy ounces (not for microbalances).
HKTAEL
The device displays the weight in taels – Hong Kong (not for microbalances).
SNGTAEL
The device displays the weight in taels – Singapore (not for microbalances).
TWNTAEL
The device displays the weight in taels – Taiwan (not for microbalances).
GRAINS
The device displays the weight in grains.
PENYWT.
The device displays the weight in pennyweights.
MILLIGR.
The device displays the weight in milligrams (not for high-capacity precision
balances).
CHINATAEL
The device displays the weight in taels – China (not for microbalances)
MOMMES
The device displays the weight in mommes.
TOLA
The device displays the weight in tolas.
BAHT
The device displays the weight in baht (not for microbalances).
MESGHAL
The device displays the weight in mesghals.
NEWTON
The device displays the weight in newtons (not for microbalances).
* Factory setting