System Settings
Parameters in the “Animal Weighing” Submenu
Parameters
Setting values
Explanation
ACTIVITY
CALM
−
Sets the intensity of the “Animal activity” to “calm”.
−
Recommended for minor movements of the sample, e.g. caused by
placement on the weighing pan.
MEDIUM*
−
Sets the intensity of the “Animal activity” to “medium”.
−
Recommended for medium movements of the sample, e.g. caused by
placement on the weighing pan.
V.ACTIVE
−
Sets the intensity of the “Animal activity” to “very active”.
−
Recommended for very active movements of the sample, e.g. caused by
placement on the weighing pan.
START
MANUAL
The “Animal weighing” application must be manually selected in the start
screen.
AUTO*
Sets the trigger to start the “Animal weighing” application to “automatic”.
* Factory setting
Parameters in the “Calculation” Submenu
Parameters
Setting values
Explanation
METHOD
MUL.*
Specifies multiplication as the method of calculation for the “Calculation”
application.
DIV.
Specifies division as the method of calculation for the “Calculation” application.
DEC.PLCS
NONE
The result of the “Calculation” application is displayed without decimal places.
1 DEC.PL.*
The result of the “Calculation” application is displayed to 1 decimal place.
2 DEC.PL.
The result of the “Calculation” application is displayed to 2 decimal places.
3 DEC.PL.
The result of the “Calculation” application is displayed to 3 decimal places.
* Factory setting
Parameters in the “Density Determination” Submenu
Parameters
Setting values
Explanation
DEC.PLCS
NONE
The result of the “Density determination” application is displayed without
decimal places.
1 DEC.PL.*
The result of the “Density determination” application is displayed to 1 decimal
place.
2 DEC.PL.
The result of the “Density determination” application is displayed to 2 decimal
places.
3 DEC.PL.
The result of the “Density determination” application is displayed to 3 decimal
places.