Save the trace before trace operation.
1108
Error
No reference marker.
Try to create the
△
marker (only for COM) before creating the reference marker
1109
Error
The data and saved track do not match with each other. The track operation is shut down.
Warning: channel settings are changed in trace operation.
1110
Error
The data and saved trace do not match with each other. Measurement conditions of the memory and data tracks should
be similar in trace operation. Try to perform trace operation on the mismatching data and memory trace.
1111
Error
The marker bandwidth is not found.
The trace conforming to the specified bandwidth standard cannot be found.
1112
Error
The peak value is not found.
The trace conforming to the specified peak standard cannot be found.
1113
Error
The target search value is not found.
The interpolation data point conforming to the search value requirements is not found.
1114
Error
One auxiliary port must be provided for reflection measurement (such as S11) to clearly identify the Port 2 measurement
of the multi-port device.
1115
Warning
As a result of measurement type or source port changes, the receiver power calibration is invalid and shut down.
1116
Warning
The non-ratio power measurement is required for receiver power calibration.
1117
Warning
The source power calibration mode matching the channel and source port is not found in the current active measurement.
Before receiver power calibration, calibrate the source power or call one source power calibration type.
1118
Error
The operation is only applicable to standard measurement.
1119
Error
The custom measurement cannot be loaded as a result of mismatch with the hardware of the network analyzer.
1120
Error
The custom measurement cannot be loaded as a result of mismatch with the software of the network analyzer.
1121
Error
The custom measurement cannot be loaded for unknown reasons.
1122
Error
Custom measurement data processing is abnormal and will be terminated. The software of the network analyzer may be
not steady. It is recommended to shut down the program and run it again
1123
Error
The operation is only applicable to custom measurement.
1124
Error
The proposed custom measurements are unavailable.
1125
Error
The proposed custom algorithm does not exist.
1126
Error
Normalization fails, as the divisor storage zone is invalid in measurement.
1127
Warning
The original data for measurement cannot be provided.
1128
Prompt
Time domain transformation or "Add” gate operation are not allowed in the selected sweep type or prohibited.
1284
Prompt
The instrument status word is changed.