Prompt
A circle cannot be properly fitted as a result of inappropriate sliding in measurement of the standard calibration kit for the
sliding load. The original impedance of the standard calibration kit is used to determine the direction of one or more
point(s).
To accurately describe the standard calibration kit, the sliding load must be moved properly to ensure sufficient sampling
along the plural circle or Smith chart. Insufficient sampling may result in inaccurate results.
599
Prompt
One unused channel is required but cannot be found. Release one channel and test it.
600
Prompt
Interpolation cannot be done to the original standard calibration kit in the segment sweep mode. Shut down the calibration
function.
601
Error
Calibration is not complete (the standard calibration kit measurement is not completed).
602
Error
The correction function is not enabled.
603
Error
The power correction function cannot be enabled. Calibrate the power meter at first.
604
Error
The sweep state is changed, the power correction data are invalid, and the power correction function is shut down.
605
Error
The power correction function is shut down, and the original sweep state is recovered.
Key parameters include the sweep type, starting frequency, frequency span and point number.
608
Error
The calibration type is not set.
Calibration can be done only when the calibration type is valid and the appropriate calibration type is selected.
609
Error
Fail to achieve the calibration characteristics.
The specified calibration kit may be one of the optional calibration kits. For example, one standard calibration is required in
response calibration, three in one-port calibration and 12 in full 2-port calibration.
610
Error
The calibration class is invalid for the current calibration type. Select a different calibration class or type.
611
Error
The standard calibration data corresponding to the selected calibration type are not found.
The original measurement data of the specified standard calibration kit is not found in the current measurement cache
zone.
612
Error
Error item data corresponding to the selected calibration type are not found.
No specified error item in the error correction cache zone (including all error coefficients of the current calibration).
613
Error
Reserved.
The access to the calibration set is not successful. The calibration set may be deleted or destroyed.
614
Error
The specified measurement is not compatible with the current calibration. Select a different measurement type or load a
different calibration type or reinstall the
calibration kit.
Prevent measurement selection so as not to shut down the measurement calibration function. Some calibration types do
not support all measurements. For example, S111-PORT calibration cannot be used in S12 calibration measurement. If
the measurement which is not supported by the current calibration is selected, prompt information will be given, and the
calibration function will be shut down.