11.2.
Measurement Methods
Following test procedures is made for S3602B. For other models, the procedures are similar.
Test of frequency accuracy
Test of frequency resolution
Test of port output harmonic suppression
Test of non-harmonic suppression of port output
Test of maximum output power of port
Test of port output power accuracy
Test of sweep range of port output power
Linearity of port output power
Test of 1dB compression level of port
Test of system trace noise
Test of dynamic range of system
Test of transmission tracking, reflection tracking, effective source matching, effective direction and effective load matching
Test of rising/falling time of pulse modulation
Test of pulse modulation switching ratio
11.2.1.
Test of Frequency Accuracy
11.2.1.1.
Items of Frequency Accuracy Test
Test the frequency accuracy indicators of frequency sweep signals output by S3602 series vector network analyzers.
11.2.1.2.
Instrument and Accessories for Test of Frequency Range and Accuracy
1) Frequency meter
HP5361B
2) RF cable
1.
Fig.11.1 Block Diagram of Frequency Accuracy Test
11.2.1.3.
Steps of Frequency Accuracy Test
1) As shown in Fig. 11.1, connect the system and preheat is for 60min at least.
2) Press [Reset] to set the initial state of S3602 series vector network analyzers.
3) Press [Stimulus] > [Sweep] > [CW] > [Frequency], and enter 10GHz. Press [Stimulus] > [Trigger] > [Hold]. The reading of the
frequency meter will be the frequency of the stimulus signal of the analyzer.
11.2.2.
Test of Frequency Resolution