![Saluki S3602 Series User Manual Download Page 295](http://html2.mh-extra.com/html/saluki/s3602-series/s3602-series_user-manual_3674132295.webp)
Tel: 886.2.2175 2930 Email: [email protected]
www.salukitec.com
295
Effective load matching;
≥48dB
≥44dB
≥44dB
Table 11.7 Transmission Tracking, Reflection Tracking, Effective Source Matching, Effective Direction and Effective Load Matching of S3602C/D
(Power setting: -5dBm; IF bandwidth: 10Hz, 23
℃
±3
℃
)
Frequency range
10MHz≤f≤2GHz
2GHz<f≤13.5GHz
13.5GHz<f≤40GHz
40GHz<f≤50GHz
Transmission tracking
±0.10dB
±0.11dB
±0.16dB
±0.20dB
Reflection tracking
±0.04dB
±0.04dB
±0.03dB
±0.04dB
Effective direction;
≥42dB
≥42dB
≥38dB
≥36dB
Effective source
matching;
≥36dB
≥31dB
≥28dB
≥27dB
Effective load
matching;
≥42dB
≥42dB
≥37dB
≥35dB
Table 11.7 Transmission Tracking, Reflection Tracking, Effective Source Matching, Effective Direction and Effective Load Matching of S3602C/D
(Power setting: -5dBm; IF bandwidth: 10Hz, 23
℃
±3
℃
)
Frequency range
10MHz≤f≤2GHz
2GHz<f≤13.5GHz
13.5GHz<f≤40GHz
40GHz<f≤67GHz
Transmission tracking
±0.10dB
±0.11dB
±0.16dB
±0.20dB
Reflection tracking
±0.05dB
±0.06dB
±0.08dB
±0.10dB
Effective direction;
≥35dB
≥41dB
≥34dB
≥32dB
Effective source
matching;
≥35dB
≥31dB
≥28dB
≥25dB
Effective load
matching;
≥35dB
≥41dB
≥33dB
≥30dB
11.1.13.
Falling Time of Pulse Modulation
Falling time of pulse modulation: ≤30ns
500MHz≤f≤50GHz
11.1.14.
Pulse Modulation Switching Ratio
Table 11.8 Indicators of Pulse Modulation Switching Ratio
Frequency range
10MHz≤f≤4GHz
4
<
f≤50GHz
Switch ratio of pulse modulation
≥64dB
≥80dB
11.1.15.
Port Damage Level
Port damage level: 26dBm or 30VDC.
Limit level of port damage: +26dBm or 30VDC. If the input is beyond the above range, the instrument may be burnt.
11.1.16.
Operation Temperature and Storage Temperature
Table 11.9 Range of Working Temperature and Storage Temperature