Operating temperature range
0°C-40
℃
Storage temperature range
-30°C - 70°C
11.1.17.
Number of Measurement Points
The number of measurement points in each frequency sweep of S3602 series vector network analyzers can be set between 1 and
32001.
11.1.18.
Measurement Display Format
S3602 series vector network analyzers have the following measurement display options: logarithmic amplitude, linear amplitude,
standing wave ratio, phase, group delay, real part, imaginary part, Smith chart, and polar coordinates.
11.1.19.
Display Window and Channel
S3602 series vector network analyzers are equipped with a 12.1-inch color LCD display, with the resolution of 1280×800. At most 16
windows and 64 channels can be set. Each window supports 8 optional channels at most. All channels can be used for display at
the same time. Each channel can be used to separately display basic S- parameters such as S11 and S21.
11.1.20.
Averaging Factor
Averaging factor range of S3602 series vector network analyzers: 1-1024.
11.1.21.
Marker Function
S3602 series vector network analyzers are configured with 9 independent marker to read and display the value of the measurement
curve. Specific functions are as follows.
a) marker search function: used for automatic search on the measurement curve, such as the maximum value, minimum value,
right peak value, left peak value, etc.
b)
△
function: used to read the difference between the current marker and reference marker.
c) Marker function: used to set the current starting value, ending value, center value, span, reference value and delay time as the
active marker value.
d) Bandwidth search function: used for automatic search to set the bandwidth (it can be set by the user, and the default value is
3dB) and calculate the Q value.
11.1.22.
Sweep Type
S3602 series vector network analyzers are configured with 5 scanning types:
a) Linear frequency: default sweep type of the instrument. The frequency linearity continuously covers the whole frequency
range.
b) Logarithmic frequency: the source frequency increases in the logarithmic step manner, and the frequency ratio of two adjacent
frequency points is identical.
c) Power sweep: used in the CW mode.
d) CW: set the single sweep frequency of the analyzer, continuously and accurately sample measurement data according to the
interval determined with the sweep time and measurement point number, and display measurement data changes over time.
e) Segment sweep: set the enabling of sweep involving multiple segments. The power level, IF bandwidth and sweep time of
each segment can be defined separately.
11.1.23.
Intermediate Frequency Bandwidth
Intermediate frequency bandwidth: min. 1Hz and max. 5MHz; stepping modes: 40 in total.
11.1.24.
Setting of Reference Level Amplitude