615
Error
A new calibration set has been created.
The newly created calibration set will be automatically named and sealed with the timestamp. Only the calibration set
subject to calibration can be saved. In case of suspension or failure in completion of calibration, the calibration set will be
deleted.
617
Error
The pre-calibration fails.
620
Error
The measurement state is changed, and the antenna and RCS calibration function is shut down.
621
Error
The measurement state is changed, and the antenna and RCS calibration function cannot be enabled.
622
Error
The calibration data are invalid, and the antenna and RCS calibration function cannot be enabled.
634
Error
Calibration set file: <x> loading failure.
The calibration set file is a collection of calibration sets and saved in the hard disc.
635
Error
Calibration set file: <x> saving failure.
File operation error and saving suspension.
636
Prompt
One calibration set is deleted.
One calibration set is successfully deleted. This may be caused as a result of application or deliberate operation of the
user.
637
Error
Calibration set file version: <x> is not compatible with the current measurement.
The calibration set is not available as a result of version nonconformity. This may be caused by software upgrading.
638
Error
Incompatible calibration set: <x>of<y>, the saved calibration set is loaded.
Some calibration sets in the calibration set file are wrong.
639
Prompt
Calibration set file: <x> is not found. A new file will be created.
The calibration set file is saved in the hard disc. After start-up of the network analyzer, the file will be searched. If the file is
found, it indicates that the file has been loaded; otherwise, a new file will be created, and the prompt will be provided.
640
Error
The specified calibration set is used at present.
It indicates that several calibration set users are performing calibration tasks, resulting in conflicts.
641
Error
The specified calibration set cannot be opened.
Many users are trying to visit the calibration set.
642
Error
Reach the maximum register number of the calibration set. Delete the old or useless calibration set before creating a new
calibration set.
The network analyzer can be configured with 100 calibration sets.
644
Error
Calibration must be effectively done once before enabling the error correction function.
This generally indicates that the calibration is not completed or there is no calibration kit matching with the selected
measurement type in the current calibration set.
646
Error
The calibration set is not loaded and the version is too new.