The new calibration file is used in the old software. The version is not compatible.
772
Error
The DSP board drive program does not work. Check the hardware. Use simulation data in the software.
Communication between the network analyzer and DSP board fails. The hardware or drive program may be faulty.
773
Error
The serial bus of the instrument does not work properly. The serial EEPROM of the instrument may be all "1" or "0". This
may be caused by hardware problems.
774
Error
Intermediate frequency overload.
848
Error
Locking failure <x>.
The source phase can't be locked properly. This may be caused by hardware failure, poor calibration or data damage in
serial EEPROM.
849
Error
Phase locking failure <x>.
850
Error
Unknown hardware error.
The communication with DSP fails as a result of hardware problems.
855
Error
The source amplitude is not stabilized.
The source amplitude cannot be properly stabilized at the required power. The indicated power is not accurate.
856
Error
The LO amplitude is not stabilized.
859
Error
YIG calibration fails.
The internal auto-calibration tuning of the YIG oscillator fails.
861
Error
The simulation slope calibration fails.
The internal simulation slope calibration fails.
865
Error
EEPROM writing fails.
The storage of calibration data in EEPROM fails. This may be caused by hardware failure.
867
Error
Try to write the input or read-only I/O port.
Try to write one input or read-only I/O port.
868
Error
Try to read the input or write-only I/O port.
Try to read the input or write-only I/O port
1025
Error
The instrument cannot be shut down normally. The program is not stable. Restart the instrument.
This is displayed when the vector network program crashes. If this message continues to appear, please consult the
service center.
1026
Error
The limit segment type is wrong.
Three limit segment types are provided: OFF. MAX and MIN.
1027
Warning
The group delay format must not be used in the CW or power sweep mode. The group delay format must not be used in
the single-frequency sweep mode; otherwise, invalid data will be generated.
1028
Warning
The limit line test fails.