Technical Indicators and Measurement Methods
This chapter describes the technical indicators and measurement methods of S3602 series vector network analyzers.
Technical indicators
Measurement methods·
11.1.
Technical Indicators
This section lists the details of performance indicators and technical parameters of S3602 series vector network analyzers. By
reading this section, the user will have an accurate understanding on main performance indicators of the product. Normal working
conditions of the instrument include the ambient temperature of 0°C to +40°C and 1 standard atmospheric pressure.
Main technical parameters of S3602 series vector network analyzers are given below. The temperature of test environment should
be 23±3°C, and the ambient temperature change must not exceed ±1°C after calibration. The pressure should be the atmospheric
pressure.
Different models and options:
S3602A/B/C/D/E are respectively used to test the frequency of 13.5GHz/26.5GHz/43.5GHz/50GHz/67GHz.
S3602A/B/C/D/E of standard configuration can be used to test the indicators of Port 1 and 2. The four-port type S3602A/B/C/D/E
can be used to test the indicators of Port 1, 2, 3 and 4.
11.1.1.
Frequency Range
Frequency range:
S3602A: 10MHz to 13.5GHz
S3602B: 10MHz to 26.5GHz
S3602C: 10MHz to 43.5GHz S3602D: 10MHz to 50GHz
S3602E: 10MHz to 67GHz
11.1.2.
Frequency Accuracy
Frequency accuracy: ±1×10
-7
(23
℃
±3
℃
)
11.1.3.
Frequency Resolution
Frequency resolution: 1Hz
11.1.4.
Port Output Harmonic Suppression
Table 11.1 Indicators of S3602A/B Port Output Harmonic Suppression
(Power setting: -5dBm, 2/3 harmonics)
Frequency range
10MHz≤f≤4GHz
4GHz<f≤13.5GHz
13.5GHz<f≤26.5GHz
Output harmonic suppression of Port 1 and 3
-51dBc
-60dBc
-60dBc
Output harmonic suppression of Port 2 and 4
-13dBc
-21dBc
-21dBc
Table 11.1 Indicators of S3602C/D Port Output Harmonic Suppression (continued)
(Power setting: -5dBm, 2/3 harmonics)