2) Setting of segment sweep type
Menu path: [Stimulus] > [Sweep] > [Sweep type].
Auxiliary menu bar: click [Segment sweep] or select [Segment sweep] in the [Setting type] dialog box.
Fig. 4.12 Setting of Segment sweep Type
Dialog box setting instructions:
a) [Independent power levels]
check box: tick it to set the independent power level of each segment. In this case, the
attenuator settings must be identical.
b) [Independent IF bandwidth]
check box: tick it to set the independent IF bandwidth of each segment.
c) [Independent sweep time]
check box: tick it to set the independent sweep time of each segment.
d) [Show table]
button: click it to create and edit the sweep table
.
e) [Hide table]
button: click it to hide the segment sweep table.
3) Segment insertion and deletion
Only the displayed segment table can be edited.
Menu path:
[Stimulus]
>
[Sweep]
>
[Segment table]
>
[Display segment table]
. The segment table sub- menu will appear.
Tick [Display segment table] in the segment table sub-menu. Click [Insert segment] in the segment table sub- menu to add a new
segment in front of the selected segment. Click [Delete segment] in the segment table sub- menu to delete the selected segment.
Click [Delete all] to delete the whole segment table.
Fig. 4.13 Segment Insertion and Deletion
a) Editing of segment table
b) Double-click the [State] box of the segment and select “ON" or “OFF" to open or close the segment.
c) Double-click the [Starting stimulus] box of the segment and enter the starting frequency of the segment.