Maintenance: Recalibrating the Scanner
American Science and Engineering, Inc.
11-12
MINI
Z
Operator Manual 342-9000-1 Rev. B ECCN EAR99
shock is applied to the system, recalibration may be required. Recalibration requires use of the Calibra-
tion Dock reference standard (see
Figure 11-9
) provided with the system.
The recalibration procedure is controlled from the tablet, using the
Calibration
menu (
Figure 11-8
),
which is accessed from the
Status Details
mode screen (see
Figure 5-10
). The button link in the
Details screen, labeled
Calibrate Scanner
, appears only for supervisor login accounts and is located
between the
Scanner
table and the
Tablet
table.
CAUTION
While the calibration procedure is not complex, the user is cautioned that it must be
performed properly and in its entirety to optimize system performance
Figure 11-8: Calibration. This link button,
Calibrate Scanner
, is on the Status Details
mode screen, but appears for supervisor user accounts only. The controls are
described in
Table 11-1
.
Table 11-1: Calibration Description
Key
Name
Description
1
Calibrate Scanner button
Appears in Status Details mode screen for supervisor user
accounts only. Tap to open Calibration screen.
2
3
1