![background image](http://html.mh-extra.com/html/rands/upv/upv_getting-started_3262652065.webp)
Startup
R&S
®
UPV
49
Getting Started 1146.2078.62 ─ 09
Internal Display Not Activated
The graphics driver of the FMR7 and FMR9 computer modules in the R&S UPV saves
the screen configuration that was last selected and detects whether an external moni-
tor is connected.
If the R&S UPV was previously operated only with an external monitor and was
switched off in that state, the internal display will remain deactivated the next time the
R&S UPV is switched on if no external monitor is connected.
Since no external monitor is detected in this case when the instrument is switched on,
the monitor output will remain deactivated even if a monitor is connected at a later
time.
An external keyboard is required to
reactivate
the screen display:
You can switch to the external monitor by using the key combination CTRL + ALT + F1
on an external keyboard.
Before March 2008:
You can switch on the internal display by using the key combina-
tion CTRL + ALT + F4 on an external keyboard.
Since March 2008
, the R&S UPV has been shipped with a new LCD that can be
switched on using the shortcut CTRL + ALT + F3.
A special feature of the R&S
UPV66 is that it has no internal display. It is therefore
possible to control an analog VGA
and
a digital DVI monitor simultaneously. The fol-
lowing settings must be performed to enable this:
Using the shortcut CTRL + ALT + F1 on an external keyboard switches on the external
VGA monitor; the shortcut CTRL + ALT + F4 switches on the external DVI monitor if it
is to be activated subsequently.
2.8.2 Changing Screen Resolution of External Monitor under Windows 7
The R&S UPV audio analyzer has a built
‑
in computer. Up to February 2007, the instru-
ments were supplied with a computer bearing the internal Rohde & Schwarz designa-
tion FMR6, from March 2007 with FMR7, from March 2012 with FMR9 and from Febru-
Connecting an External Monitor