Getting Started
R&S
®
UPV
100
Getting Started 1146.2078.62 ─ 09
rotary-knob click, and then highlighting and activating "Preset (Load Default)" in the
displayed list in the same way.
2. Step: Setting the required generator signal
a) The basic settings for configuring the generator are made in the "Generator
Config" panel.
The panel must have the focus before it can be operated. This is already the
case with the default setup, otherwise press the WINDOW </> keys until the
color of the title bar in the desired panel changes to blue.
The frequency response is usually measured on both channels; this is already
configured in the "Channel" field.
Depending on the DUT, "Output Type" can be switched to balanced circuitry;
various output impedances can then be selected in the field below.
If necessary, the required parameter field can be highlighted with the rotary
knob; this is then indicated by a selection frame (see image below). A rotary-
knob click toggles between the two radio buttons.
b) Floating output circuitry for the generator (see
on page 80) and a maximum signal frequency of 22
kHz are generally
the correct settings to be used.
The following settings are also suitable for the majority of applications and do
not have to be adjusted unless necessary.
This completes the settings in the "Generator
Config" panel.
Introduction to Instrument Operation Using Examples