![background image](http://html.mh-extra.com/html/rands/upv/upv_getting-started_3262652031.webp)
Startup
R&S
®
UPV
15
Getting Started 1146.2078.62 ─ 09
●
When the WINBAR key is pressed, the WINBAR containing the names of all
opened panels or graphical windows is displayed in place of the softkey bar. Press-
ing the associated softkey puts the focus on the corresponding panel or graphical
window, i.e. it becomes active. At the same time, the system returns to normal
operating mode (see above).
MODIFY
Opens a selection of functions on the softkey bar; these functions can be used in con-
junction with the rotary knob to move or resize the active panel or graphical window
HIDE
Minimizes ("hides") the active panel or graphical window
CLOSE
Closes the active panel or graphical window
HELP
Pressing this key displays a context-sensitive help text
SAVE
Opens a window for storing instrument settings
LOAD
Opens a window for loading predefined instrument settings
PRESET
Loads the
factory default settings
of the instrument or carries out a
Master PRESET
.
●
Factory setting:
Pressing the button briefly displays the security prompt "Do you want to preset the
UPV?". If you answer this question with "Yes", the factory default settings are
restored immediately.
●
Master PRESET:
With the Master PRESET, you can easily get out of an unsuitable screen and mon-
itor configuration and back to an operable instrument: Switch the instrument off and
on again. During the software start, the OUTPUT Off-LED is illuminated briefly in
yellow. As soon as this LED is illuminated in yellow, press the PRESET button
briefly. The LED flashes for confirmation. Within the next 5 seconds, press the
PRESET key again briefly. The Master PRESET is now carried out and forces a
switchover to the external monitor with a screen resolution of 800 x 600 pixels and
the factory default settings.
2.1.11 Analog Audio Interfaces
Output and input sockets of the analog generators and analyzers
Description of Front View