![background image](http://html.mh-extra.com/html/rands/upv/upv_getting-started_3262652048.webp)
Startup
R&S
®
UPV
32
Getting Started 1146.2078.62 ─ 09
3. Press the POWER switch on the front of the instrument.
The instrument is disconnected from the AC supply, and power is removed from all
modules.
For remote control, you can also shut down the operating system by using the
SYSTem:SHUTdown
SCPI command.
Alternatively, the operating system can also be shut down in the same way as other
Windows applications by using the mouse: Start, Shut Down, OK or the red Shutdown
icon.
2.4 Function Check
The audio analyzer monitors the most important instrument functions automatically
when it is switched on.
2.5 Presets
Pressing the PRESET key loads a set of defined instrument settings.
The preset function sets all parameters and settings, even those of inactive operating
modes.
The most important parameters set here are:
Generator frequency (analog)
1 kHz
Generator level (analog)
0.1 V
Generator signal
Sinusoidal
Generator frequency (digital)
997 Hz
Generator level (digital)
0.1 FS
Generator signal
Sinusoidal
Generator output impedance
5
Ω
Analyzer interface (analog)
XLR measuring input (analog)
Analyzer interface (digital)
XLR measuring input (digital)
Analyzer function
RMS
Function Check