Getting Started
R&S
®
UPV
112
Getting Started 1146.2078.62 ─ 09
a) The easiest way to load the factory settings of the instrument is to press the
PRESET key on the front panel. Alternatively, the default setup can also be
loaded via the menu bar.
2. Step: Setting the required generator signal
a) In the "Generator
Config" panel, the basic settings for configuring the generator
in this example are the same as those set for the frequency response measure-
ment described in the previous section.
Depending on the DUT, "Output
Type" may have to be switched to balanced
circuitry.
b) Next, the "Generator
Function" panel is activated.
Sine-wave signals are required. This setting is one of the presets.
The total harmonic distortion is often only specified for the reference frequency
1
kHz (for analog applications). Fixed frequencies are therefore used for mea-
surement; as a result, "Sweep
Ctrl" must remain set to Off.
c) The frequency 1 kHz is already preset in the default instrument setup.
Only the output level required for the measurement may have to be adjusted (in
the Voltage field).
The remaining fields are not relevant for this example. Configuration of the gen-
erator is therefore complete.
Introduction to Instrument Operation Using Examples