In the default instrument setup, the numeric measurement display is also already
open. If the DUT is correctly connected to the audio analyzer, measured values will
already be displayed.
The panels for configuring the measured value display are used to determine how
the measurement results are presented (i.e. which unit, how many digits, etc.). A
series of panels for the different measurement windows are provided for this pur-
pose.
The "Function
Config" panel is shown here as an example; the first column of the
numeric measurement display is configured in this panel, i.e. the display fields
which show the measured level values for the frequency response measurement
described here.
The panel has two columns; to simplify matters, only Channel
1 is shown here.
The unit for the numeric measurement displays is set in the Unit field.
The Engineering parameter is used to determine whether the display automatically
toggles between µV, mV and V, for example, or whether other unit formats are
desired.
If the measurement results are to be displayed as relative values, they are referred
to the reference value set in the Reference field.
Other parameter fields in this panel are used to set limits and to display extreme
values; see also
on page 121 as well as the oper-
ating manual on the documentation CD.
5. Step: Starting the sweep
Although the analyzer is already supplying measurement results, the status mes-
sage "Sweep Waiting" above the softkeys indicates that the sweep has not yet
been started although it has been configured in the "Generator Function" panel.
Introduction to Instrument Operation Using Examples