R&S AFQ100A
General Instrument Settings
1401.3084.32 4.97
E-3
GPIB - Setup Remote
The
Remote Channel Settings
dialog provides access to the GPIB and Ethernet settings. The
GPIB
dialog can be selected in the
Remote
section of the
Setup
menu.
GPIB channel address
Sets the IEC/IEEE bus address of the instrument.
Remote-control command
:
SYST:COMM:GPIB:ADDR 28
Visa Resource String
Indicates the visa resource string. This string is used for remote
control of the instrument. A separate string is provided for remote
control via the LAN and the USB interface
Remote-control commands:
SYST:COMM:USB:RES?
Response:
USB::0x0031::100001::INSTR
Local
Switches the instrument to operate in local control mode.
Switching from remote to local control mode can be also done with
one of the following actions:
with the interface command >L via the remote-control interface
with the key combination
CTRL + Q
Remote-control command: n.a.