R&S AFQ100A
AFQ Commands - Setup Section
1401.3084.32
6.72
E-3
AFQ Commands - Setup Section
Calibration Subsystem
The Calibration system contains the commands for adjustment. Adjustment is triggered by the query
commands. The response "0" indicates error-free adjustment, and the response "1" means that an error
occurred during adjustment.
Command
Parameters
Default
unit
Remark
CALibration:ALL[:MEASure]?
Query only
CALibration:AOUTput<CH>:AMPLitude?
Query only
CALibration:AOUTput<CH>:DELay?
CALibration:ALL[:MEASure]?
The command starts all internal adjustments for which no external measuring equipment is needed.
Example:
"
CAL:ALL:MEAS?"
'starts the adjustment of all functions for the entire instrument.
Response: "0"
'adjustment has been performed successfully.
*RST value
Resolution
Options
SCPI
-
-
Compliant
CALibration:AOUTput<CH>:AMPLitude?
The command adjusts the amplitude of the (selected) analog output.
The instrument generates a series of signal levels and converts them using a precise ADC. The
procedure takes about 30 seconds.
The command is a query command and therefore has no *RST value.
Example:
CAL:AOUT:AMPL?
'starts all internal adjustments for which no external measuring equipment is
needed. The adjustments with external measuring equipment are described in
the
service manual
.
Response: '0'
'calibration was successful
*RST value
Resolution
Options
SCPI
-
-