R&S AFQ100A
AFQ Commands - Signal Generation
1401.3084.32 6.51
E-3
[SOURce:]EQUalizer:IQ:Q:FREQuency
0 ... 1 GHz
This command enters frequency values into the current data table of the Q path.
The command triggers an event and therefore has no query form and no *RST value.
Example:
EQU:IQ:Q:FREQ 10000,20000,30000
'enters three frequency values into the current data table.
*RST value
Resolution
Options
SCPI
-
1 Hz
Device-specific
[SOURce:]EQUalizer:IQ:Q:GAIN
-5 ... +50 dB
This command enters gain values into the current data table of the Q path.
The command triggers an event and therefore has no query form and no *RST value.
Example:
EQU:IQ:Q:GAIN 0,-5,-3
'enters three gain values into the current data table.
*RST value
Resolution
Options
SCPI
-
0.001 dB
Device-specific
[SOURce:]EQUalizer:IQ:Q:PHASe
-360 ... +360 deg
This command enters phase values into the current data table of the Q path.
The command triggers an event and therefore has no query form and no *RST value.
Example:
EQU:IQ:Q:PHAS 0,-5,-10
'enters three gain values into the current data table.
*RST value
Resolution
Options
SCPI
-
0.01 deg
Device-specific
[SOURce:]EQUalizer:IQ:Q:POINts?
This command queries the number of data points of the current data table of the I path.
The command is a query command and therefore has no *RST value.
Example:
EQU:IQ:I:SEL "D:/TEMP/TESTPREDIQ"
'selects the data table.
EQU:IQ:Q:POIN?
'queries the number of data points of the current data table.
*RST value
Resolution
Options
SCPI
-
-
Device-specific