R&S AFQ100A
Alphabetical List of SCPI-Error Messages
1401.3084.32 9.8
E-3
Self test failed ... (-330)
An error was detected in the selftest named after the semicolon. An error-free operation of the module
concerned is no longer guaranteed.
SCPI:
Device specific error - sets bit 3 in the ESR register.
Settings conflict ... (-221)
There is a setting conflict between the two parameters indicated after the semicolon.
Example:
The set FSK deviation is too large for the selected symbol rate.
Remedy:
One of the given values has to be corrected to obtain a valid output signal.
SCPI:
Execution error - sets bit 4 in the ESR register.
String data not allowed (-158)
The command contains a legal string data element which is not allowed at this point.
Example:
A text parameter is set in quotation marks -
SOURce:FREQuency:MODE
"FIXed"
.
SCPI:
Command error - sets bit 5 in the ESR register.
Suffix not allowed (-138)
A suffix is not allowed for this command or at this point of the command.
Example:
Command *RCL does not permit indicating a suffix.
SCPI:
Command error - sets bit 5 in the ESR register.
Suffix too long (-134)
The suffix contains more than 12 characters.
SCPI:
Command error - sets bit 5 in the ESR register.