R&S AFQ100A
AFQ Commands - Signal Generation
1401.3084.32 6.43
E-3
[SOURce:]ROSCillator:EXTernal:SBANdwidth
WIDE | NARRow
Reference clock
external
only!
The command sets the synchronization bandwidth, i.e. the target range of the synchronization
circuit.
Parameters:
NARRow
A narrow synchronization bandwidth is selected, the synchronization circuit
works slowly but with less jitter.
WIDE
A wide synchronization bandwidth is selected, the synchronization circuit
works fast.
Example:
ROSC:EXT:SBAN WIDE
'sets a wide synchronization bandwidth.
*RST value
Resolution
Options
SCPI
NARRow -
Device-specific
[SOURce:]ROSCillator:SOURce
INTernal | EXTernal
The command sets an internal or external reference oscillator source.
Note:
The
Internal Reference Clock Source
is available only in the reference clock synchronization
modes
Normal
and
Master
. If the instrument is set to
Slave
or
Last slave
mode the reference clock
source is set to
External
.
Example:
ROSC:SOUR EXT
'selects an external reference oscillator source.
*RST value
Resolution
Options
SCPI
INTernal -
Device-specific