R&S AFQ100A
Baseband Block
1401.3084.32 4.9
E-3
Sample Clock - Table
Sine Test Signal
Enters the sample clock frequency for generating a table sine test
signal.
1 kHz ... 300 MHz
The sample clock rate must not exceed the maximum ARB clock rate
of 300 MHz. Sample clock rates between 300 MHz and 600 MHz are
not available and set to maximum ARB clock rate of 300 MHz.
Remote-control command:
SOUR:TSIG:CLOC 300e6
Table Length - Table
Sine Test Signal
Enters the number of sample values of a waveform.
1 kHz ... 300 MHz
The sample clock rate must not exceed the maximum ARB clock rate
of 300 MHz. Sample clock rates between 300 MHz and 600 MHz are
not available and set to maximum ARB clock rate of 300 MHz.
The number of sample values is automatically restricted by reference
to the set frequency.
Remote-control command:
SOUR:TSIG:TSIN:TLEN
Offset - Table Sine
Test Signal
Enters the DC component of the resulting signal in % or FS units. The
abbreviation FS means full scale.
Remote-control command:
SOUR:TSIG:TSIN:OFFS